A recall has been issued for the Covidien Medi-Trace Cadence and Kendall
 Defibrillation Electrodes because the electrodes will not work with 
Philips FR3 and FRx Automated External Defibrillators.  
Philips made 
changes to the design of their connectors for the FR3 and FRx AEDs, and 
because of these changes, the Covidien defibrillation electrodes will 
not work with these AEDs.  Use of these electrodes could cause a delay 
in delivering electrical therapy that is needed to revive patients. 
Additionally, their use could lead to severe health problems such as 
stroke, organ failure, or death if alternative defibrillator electrodes 
are not readily available. Covidien received two reports of injuries. 
There was a delay in resuscitating one patient which may have 
contributed to the subsequent death of the patient.  
For more 
information, please see: http://www.fda.gov/
 